ASTM D5386-24 PDF
1.1This test method covers an instrumental method for the CIE (Commission International de l”Eclairage) tristimulus measurement of the color of near-clear liquid samples. The transmission measurement values are converted to the color scale in the platinum-cobalt (Pt-Co) system.
1.2This test method has been found applicable to the color measurement of clear, liquid samples, free of haze, with nominal Pt-Co color values in the 0 to 100 range. It is applicable to non-fluorescent liquids with light absorption characteristics similar to those of the Pt-Co color standard solutions.
1.3The tristimulus colorimetry may be calculated from spectral measurements or be obtained directly from a colorimeter. Practice E1455 provides some guidance on colorimeters, while Test Methods D1686 and D2108 deal with the visual and instrumental measurement of near-clear liquids.
1.4In determining the conformance of the test results using this method to applicable specifications, results shall be rounded in accordance with the rounding off methods of Practice E29.
1.5The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.