1.1This test method covers the analysis of drinking water. It is also applicable to many environmental and waste waters when adequate validation is included.
1.2This test method covers the determination of halomethanes, haloethanes, and some related extractable organohalides amenable to gas chromatographic measurement. The applicable concentration range for trihalomethanes is from 1 μg/L to 200 μg/L. Detection limits depend on the compound, matrix, and on the characteristics of the gas chromatographic system.
1.3For compounds not specifically included in the precision and bias section the analyst should validate the test method by collecting precision and bias data on actual samples.
1.4Confirmation of component identities is obtained by observing retention times using gas chromatographic columns of different polarities. When concentrations are sufficiently high (>50 μg/L) confirmation with halogen specific detectors or gas chromatography/mass spectrometry (GC/MS) may be used. Confirmation of purgeable compounds at levels down to 1 μg/L can be obtained using Test Method D3871 with GC/MS detection.
1.5The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 8.
1.7This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.